AMETEK ReFlex User Manual

Page 139

Advertising
background image

AC Power Supplies

ReFlex Power™

4-54

M380056-01 Rev K

4.6.8

V

OLTAGE

ADC

O

FFSET

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

L

OW

The following procedure calibrates offset of the voltage ADC for the voltage
range low.

1. Ensure that no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST<n>

3. Perform the voltage ADC offset calibration

CAL<n>:MEAS:VADC:OFFSET:CALC

4. Send the system error query command to make sure calibration was successful.

SYST:ERR?

5. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK

6. Optional Commands:

To query the voltage ADC offset:

CAL<n>:MEAS:VADC:OFFSET ?

7. Reset the unit

*RST<n>