AMETEK ReFlex User Manual

Page 165

Advertising
background image

AC Power Supplies

ReFlex Power™

4-80

M380056-01 Rev K

14. Set the output frequency to 250Hz

SOUR<n>:FREQ 250

15. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR2 <value-2>

16. Set the output frequency to 440Hz

SOUR<n>:FREQ 440

17. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR3 <value-3>

18. Set the output frequency to 630Hz

SOUR<n>:FREQ 630

19. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR4 <value-4>

20. Set the output frequency to 820Hz

SOUR<n>:FREQ 820

21. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR5 <value-5>

22. Set the output frequency to 1010Hz

SOUR<n>:FREQ 1010

23. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR6 <value-6>

24. Set the output frequency to 1200Hz

SOUR<n>:FREQ 1200

25. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL<n>:OUTP:FREQ:CURR7 <value-7>

26. Perform the voltage calibration using following command