Tektronix CSA8000B User Manual

Page 108

Advertising
background image

Acquiring Waveforms

3- 34

CSA8000B & TDS8000B User Manual

Overview

Control elements and resources

To acquire in FrameScan mode (cont.)

Set the bit rate 8.

Set the horizontal scale so one acquisition record is
equal to one bit. Use one of the two methods that follow:

H Automatic: If your signal to be scanned matches a

communications standard, select it from the Comm
Standard list. Choosing a standard sets the bit rate
and start bit; otherwise, if you know the bit rate, you
can set the bit rate manually using the Bit Rate
box.

H Manual: Adjust the Scale control to a setting that

results in a display of both edges of the bit. For
example, setting 1/8 of a bit per division (0.125
bits/div) yields 1 bit in 8 divisions, which fits nicely
on screen.

Select a comm.

standard, or...

Set the bit rate

manually

Set to 1/8 bit

per division

Set the starting

horizontal posi-

tion

9.

Set the initial horizontal position to the first bit you want
to acquire. Use one of the two methods that follow:

H Automatic: Enter your desired start bit location, and

then check the Auto Position box to enable the
instrument to set the position as near as possible to
match the bit specified in the Start Bit box.

H Manual: Adjust the Position control to align the

start of a bit to desired location in the frame.

Tip. The latter method is useful when you need to
manually align a bit or waveform to a mask on the
display.

Or set manually

First set the

start bit, ...

and then enable

Auto Position

Enable

FrameScan

10. In the dialog box, click to check the FrameScan

Enabled box. See right.

11. To restart the scan at the first bit at any time, click the

Reset button.

Click to

restart scan

Check to

start scan