Tektronix CSA8000B User Manual

Page 48

Advertising
background image

Incoming Inspection

1- 30

CSA8000B & TDS8000B User Manual

NOTE. At some temperatures, there may be extraneous data points after the first
half cycle when viewing the front-panel Internal Clock output (as is done in this
step). This behavior may also occur when viewing multiple cycles in TDR mode.
In both cases, this behavior is normal.

H Rotating the Horizontal SCALE knob clockwise expands the waveform

on-screen (more horizontal divisions per waveform period), counter-
clockwise rotation contracts it, and returning the horizontal scale to
1 s/div

returns the period to about five divisions. Leave the time base

set to 1 s/div.

H The horizontal POSITION knob positions the signal left and right

on-screen when rotated.

NOTE. The signal will not move past the minimum position setting.

Control bar

Vertical scale

setting

Horizontal

scale setting

Internal clock

signal

s

Figure 1- 13: Main time base verification