Measurement reference parameters and methods, All sources – Tektronix CSA8000B User Manual

Page 318

Advertising
background image

B- 56

CSA8000B & TDS8000B User Manual

Measurement Reference Parameters and Methods

This reference topic describes the reference parameters (levels and crossings)
used in taking the measurements.

All Sources

The methods available for calculating reference levels used in taking automatic
measurement follow. The methods are shown using a pulse, but they also apply
to RZ and NRZ waveforms.

Reference level calculation methods

High (50 mV)

High reference

Mid reference (0 mV)

Low reference

Low (--50 mV)

90%

10%

90 mV

10 mV

--40 mV

40 mV

90 mV

10 mV

50%

50 mV

50 mV

0 mV

1. Relative Reference is calculated as percentage of the High/Low range.

2. High Delta Reference is calculated as the absolute values from the High Level.

3. Low Delta Reference is calculated as absolute values from the Low Level.

4. Absolute Reference is set by absolute values in user units.

5. AOP (not shown) measures the Average Optical Power of the waveform and uses it
as the Mid Ref level. See Pulse Crossings and Mid-reference Level AOP on page
B--58 for more information.

Figure B- 1: Reference-level calculation methods

Reference-Level

Calculation Methods