Incoming Inspection

1- 20

CSA8000B & TDS8000B User Manual

Perform the Compensation

This procedure uses internal routines to verify that the instrument compensates
properly.

Equipment required

For sampling modules:

H 50 Ω terminations on all electrical module channels (Tektronix

part number 015-1022-xx).

H Dust covers on all optical module channels.

The sampling modules ship from Tektronix with the proper termina-
tions and dust covers installed.

Prerequisites

First, all sampling modules to be compensated must be installed as
outlined in their user manuals.

Second, power on the instrument and allow a 20 minute warm-up
before doing this procedure.

1.

Run the compensation routines:

a.

From the application menu bar, select Utilities, and then select Com-
pensation.

In the Compensation dialog box, the main instrument (mainframe) and
sampling modules are listed. The temperature change from the last
compensation is also listed. See Figure 1--6.

Click to select compensate

Click to start compensation

Choose all as targets

Figure 1- 6: Compensation dialog box