Appendix B: Automatic Measurements Reference

B- 14

CSA8000B & TDS8000B User Manual

Pulse Measurement - Area

Table B--3 describes each pulse measurement in the area category. See Table B--1
on page B--2 amplitude-category measurements; see Table B--2 on page B--8
for timing-category measurements.

Table B- 3: Pulse Measurements - Area

Name

Definition

Area

The area under the curve for the waveform within the measurement region. Area measured
above ground is positive; area measured below ground is negative.

over the measurement region

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a Measurement on page 3--83.

For best results, optimize the vertical resolution before taking this measurement. See To
Optimize the Vertical Resolution on page B--69.

Area

=

waveform(t) dt,

Cycle Area

The area under the curve for the first waveform period within the measurement region. Area
measured above ground is positive; area measured below ground is negative.

over the first waveform period within the
measurement region.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).

For best results, optimize the vertical resolution before taking this measurement. See To
Optimize the Vertical Resolution on page B--69.

Area

=

waveform(t) dt,