Measuring Waveforms

3- 78

CSA8000B & TDS8000B User Manual

High (mode)

Low (mode)

Low (min/max)

High (min/max)

High (mean)

Low (mean)

Mid reference

Figure 3- 20: High/Low tracking methods

H Mean (of Histogram) sets the values statistically. Using a histogram, it

selects the mean or average value derived using all values either above or
below the midpoint (depending on whether it is defining the high or low
reference level). This setting is best for examining eye patterns and optical
signals. See Figure 3--20.

H Min-max uses the highest and lowest values of the waveform record. This

setting is best for examining waveforms that have no large, flat portions at a
common value, such as sine waves and triangle waves — almost any waveform
except for pulses. See Figure 3--20.

H Auto switches between methods. Auto method first attempts to calculate the

high and low values using the Mode method. Then, if the histogram does not
show obvious consistent high and low levels, Auto method automatically
switches to the Min/Max or Mean method.

For example, the Mode histogram operating on a triangle wave would not
find consistent high and low levels, so the instrument would switch to
the Min/Max mode. Consistent high and low levels would be found on a
square wave, so the Auto mode would use the Mode method.

When setting High/Low method, be aware of these operating behaviors:

H The tracking settings are not global; that is, you can independently set the

method used for each of Meas 1 -- Meas 8.

H You can turn off tracking for either or both the High and Low levels and

enter them directly.

H Not all tracking methods are appropriate for all measurements. If you cannot

set the tracking method, the controls will be disabled.