Tektronix CSA8000B User Manual

Page 111

Advertising
background image

Acquiring Waveforms

CSA8000B & TDS8000B User Manual

3- 37

Overview

Control elements and resources

To catch a bit error (cont.)

Set conditional

acquisition and

start testing

6.

From the application menu bar, select Setup, and then
select Acquire.

7.

In the Acq Setup dialog box (see right), check the
Condition option under Stop After.

8.

In the Condition pulldown list, select Mask Total Hits
and set a count of one in the count box. These settings
will stop acquisition on a violation of any of the masked
areas on screen. See below.

For more

information

9.

For more help on using FrameScan acquisitions, click
the Help button in the Horz Setup dialog box to
display contextual help on screen.

See page 3--167 to learn about using online help.

End of Procedure

Advertising
This manual is related to the following products: