Using Masks, Histograms, and Waveform Databases

CSA8000B & TDS8000B User Manual

3- 145

H Masks are saved with setups, so you can save sets of masks by defining

them, and then storing the instrument setup. Displayed masks are overwrit-
ten when you recall a stored setup, select a standard mask, or initialize the
instrument.

Use the procedure that follow to set up the instrument to mask test a waveform
against a mask standard or user-defined mask set.

Overview

To mask test a waveform

Related control elements & resources

Prerequisites 1.

The instrument must have at least one waveform turned
on.

See Displaying Waveforms on page 3--53 for

information on displaying waveforms.

Access the Mask
Setup dialog box

2.

Select Mask from the Setup menu to display the Mask
Setup dialog box.

To Mask Test a Waveform