CSA8000B & TDS8000B User Manual

Overview

Related control elements and resources

To take automatic measurements (cont.)

To measure a

database

8.

From the application menu bar, select Setup, and then
select Measurement. See right.

9.

In the Meas Setup dialog box, make sure the
measurement (one of Meas1 through Meas8) is
selected.

10. In the Source tab, check the Use Wfm Database option

as shown below.

Tip. If, at the time you first create a measurement, the
measurement source you select is displayed as a
waveform database, the database will automatically be
measured. Uncheck the User Wfm Database option if
you want to measure the waveform instead of the
database.

For more in-

formation

11. Press the Help button in the Meas Setup dialog box to

access the online help.

12. See Appendix B: Automatic Measurements Reference,

on page B--1 for a list of the measurements and their
definitions.

See page 3--167 to learn about using online help.

End of Procedure