Incoming Inspection

CSA8000B & TDS8000B User Manual

1- 33

3.

Hook up the signal source: Connect the SMA cable from the Internal Clock
output through a 10x attenuator to 80E00 sampling module input channel 3
as shown in Figure 1--16. Connect BNC cable to External Gate input at rear
panel.

CSA8000/TDS8000

SMA cable from

INTERNAL CLOCK

output to 80E00 C3 input

10x attenuator

BNC cable attached to TRIGGER

GATE (TTL) on the rear panel.

TRIGGER

GATE (TTL)

Rear panel

Figure 1- 16: Hookup for the gated trigger tests

4.

Set up the instrument:

a.

Push the Trigger MENU front-panel button to display the Trig Setup
dialog box.

b.

Click Internal Clock under Trigger Source in the Trig Setup dialog
box. The Internal Clock rate should be set to 200kHz.

c.

Verify that the Gated Trigger option in Enhanced Triggering section is
selected (check box is checked). See To Use Gated Trigger, step 4 on
page 3--51.

d.

Turn the Vertical SCALE knob to set the vertical scale to 50 mV/div.
The channel scale readout is displayed in the Control bar at the bottom
of the graticule.

5.

Set the time base: Set the Horizontal SCALE to 2 s/div. The horizontal
scale readout is displayed in the Control bar at the bottom of the graticule.

6.

Set the display for Normal and Show Vectors (enable). See To Set Display
Styles on page 3--68.