Incoming Inspection

1- 34

CSA8000B & TDS8000B User Manual

7.

Push the Horizontal MENU button, the Mode in All Timebases must be set
to Lock to Int. 10 MHz.

8.

Verify that Triggering occurs: Verify signal is triggered with waveform
on-screen. See Figure 1--17 on page 1--34.

Control

bar

Vertical scale

setting

Horizontal

scale setting

Internal

clock

signal

Triggered signal indicator

Figure 1- 17: Signal triggered

9.

Disable trigger: Install 50

Ω terminator cap to the end of the cable that is

attached to the rear-panel gated trigger BNC. See Figure 1--16 on page 1--33.

10.

Verify that the Gated Trigger functions: Verify signal is not triggered (gate
disabled). Signal freezes on the screen above to indicate triggering has
stopped. See Figure 1--18 on page 1--35. Note the Not Trigd indication at the
top of the window.

a.

Push the CLEAR DATA button.

b.

Verify signal is not triggered with no waveform on-screen (see Fig-
ure 1--19 on page 1--36). Note the Not Trigd indication at the top of the
window.