Appendix B: Automatic Measurements Reference

CSA8000B & TDS8000B User Manual

B- 23

Table B- 4: RZ Measurements - Amplitude (cont.)

Name

Definition

RZ Peak-to-Peak Noise

The maximum range of the data distribution sampled within a fixed width vertical slice located at
the center of the Eye Aperture at the High or Low levels. See RZ Eye Aperture Parameters
on B--62.

PkPk noise = Highpp

or

PkPk noise = Lowpp

The Eye Aperture is adjustable and defaults to 5% of the RZ pulse width. The High or Low
selection for Noise At control in the Measurement Setup dialog specifies that the measurement
is to be performed on the logical 1 or 0 levels.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.

For best results with this measurement:

H Perform a Dark Level compensation before taking this measurement if the source of the

measured waveform is an optical channel. See To Perform Dark-Level and User
Wavelength Gain Compensations on page 3--98.

H Perform Autoset or otherwise optimize the vertical resolution before this measurement, i.e.,

increase the overall vertical size of the waveform (but without producing off-screen
waveform points). See How to Optimize the Vertical Resolution on page B--70.