Tektronix CSA8000B User Manual

Page 124

Advertising
background image

Triggering

3- 50

CSA8000B & TDS8000B User Manual

Use the procedure that follows when setting up the instrument to use the gated
trigger. Gated trigger is only available with Option GT installed.

Overview

To use gated trigger

Control elements and resources

Prerequisites 1.

The Acquisition system should be set to Run, and the
vertical and horizontal controls should be set appropri-
ately for the signal to be acquired.

2.

Trigger on your input signal. Use the procedure To
Trigger on page 3--48 as needed.

Note that you must supply the input signal and the TTL
gating signal to the appropriate instrument inputs. The
instrument does not control or generate these signals.

See Sampling Module User Manuals for sampling module

installation. See page 3--24 of this manual for acquisition

setup.

Access the

Setup trigger

dialog box

3.

From the application menu bar, select Setup, and then
select Trigger. See right. You also can select this menu
by pushing the Trigger MENU button.

To Use Gated Trigger