Tektronix CSA8000B User Manual

Page 315

Advertising
background image

Appendix B: Automatic Measurements Reference

CSA8000B & TDS8000B User Manual

B- 53

Table B- 8: NRZ Measurements - Timing (cont.)

Name

Definition

NRZ Phase

Where:

H Tcross1 of source1 is the time of the first crossing of either polarity on source 1.
H Tcross3 of source1 is the time of next crossing on source 1of the same polarity as

Tcross1.

H Tcross1 of source2 is the time of the first crossing of either polarity on source 2 after

Tcross1 of source1.

H All Tcrossings are at the mid-reference level, which is adjustable and defaults to 50%

of the NRZ eye amplitude. See NRZ Crossings on page B--64.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a Measurement on page 3--83.

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See Use a Waveform Database on page B--70.

Phase

= Tcross1 of source2 − Tcross1 of source1

Tcross3 of source1

− Tcross1 of source1 ⋅

360

NRZ Pk-Pk Jitter

The delta between the minimum and maximum of time crossings, with the mean of the
histogram being Tcross.

Pk-PK Jitter = Tcrosspp

The Jitter At control in the Measurement Setup dialog specifies whether the jitter is to be
measured at the eye cross or at the mid-reference level. See To Localize a Measurement on
page 3--83.

The mid-reference level is adjustable and defaults to 50% of the eye amplitude.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.

Advertising
This manual is related to the following products: