Appendix B: Automatic Measurements Reference

B- 34

CSA8000B & TDS8000B User Manual

Table B- 5: RZ Measurements - Timing (cont.)

Name

Definition

RZ Pulse Symmetry

RZ Pulse Symmetry measures to what extent the RZ pulse is symmetrical around the peak at
the mid-reference level. The pulse peak is the center of the interval, sized to Eye Aperture,
which yields the maximum mean vertical value. See RZ Eye-Aperture Parameters on page
B--62.

Where:

H Tcross1 and Tcross2RZFig_CrossLevels>Second are the time crossings of the RZ

pulse of the mid-reference level. See RZ Crossings on page B--61.

H Tcross0 is the time coordinate of the pulse peak.

The mid-reference level is adjustable and defaults to 50% of the RZ maximum pulse amplitude.
See Mid-reference level on page B--69.

If gating is enabled, the measurement region is constrained between the Start Gate (G1) and
Stop Gate (G2). See To Localize a Measurement on page 3--83.

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See Use a Waveform Database on page B--70.

RZ Pulse Symmetry [%]

= 100 ×

(Tcross0

− Tcross1)

(Tcross2

− Tcross1)

RZ Rise Time

RZ Rise Time characterizes the positive slope of the RZ pulse by computing the time interval
between the mean crossings of the low reference level and the high reference level.

RZ Rise Time = TcrossH -- HcrossL

Where:

H TcrossH is the mean of the histogram of the crossing of the high reference level
H TcrossL is the mean of the histogram of the crossing of the low reference level. See

RZ Measurement Reference Levels on page B--60.

The adjustable low reference and high reference levels default to 20% and 80% of the RZ
maximum pulse amplitude.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.