Appendix B: Automatic Measurements Reference

CSA8000B & TDS8000B User Manual

B- 13

Table B- 2: Pulse Measurements - Timing (cont.)

Name

Definition

+Width

The horizontal interval between the crossings of the rising and falling edges at the mid-refer-
ence level of the first positive pulse in the measurement region.

+Width = Tcross2 – Tcross1

Where Tcross1 and Tcross2 are the two consecutive horizontal crossings on the first
positive pulse.

The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.

See Measurement Reference Parameters and Methods on page B--56 or in the online help. Also
see Reference Levels Method on page 3--79.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a Measurement on page 3--83.

- Width

The horizontal interval between the crossings of the falling and rising edges at the mid-refer-
ence level of the first negative pulse in the measurement region.

--Width = Tcross2 – Tcross1

Where Tcross2 and Tcross1 are the two consecutive horizontal crossings on the first
negative pulse.

The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).