Appendix B: Automatic Measurements Reference

CSA8000B & TDS8000B User Manual

B- 15

Return-to-Zero (RZ) Measurements - Amplitude

Table B--4 describes each RZ measurement in the amplitude category. See Table
on B--5 on page B--29 for timing category measurements; see Table B--6 on page
B--36 for area category measurements.

Table B- 4: RZ Measurements - Amplitude

Name

Definition

RZ AC RMS

The root mean square amplitude, minus the DC component, of the waveform that is sampled
within the measurement region.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a Measurement on page 3--83.

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See Use a Waveform Database on page B--70.

For best results with this measurement:

H Perform a Dark Level compensation before taking this measurement if the source of the

measured waveform is an optical channel. See To Perform Dark-Level and User
Wavelength Gain Compensations on page 3--98.

H Optimize the vertical resolution before taking this measurement. See To Optimize the

Vertical Resolution on page B--69.

RZ Amplitude

The difference between the logical 1 level (High) and the logical 0 level (Low) of the RZ signal.
Both High and Low levels are measured within the Eye Aperture.

RZ Amplitude = High – Low

Where High and Low are the logical 1 and 0 levels.

The Eye Aperture is adjustable and defaults to 5% of the RZ pulse width.

See RZ Eye Aperture Parameters on B--62.

If enabled, measurement gates, constrain the measurement region to the area between the
Start Gate (G1) and Stop Gate (G2).

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.

For best results with this measurement:

H Perform a Dark Level compensation before taking this measurement if the source of the

measured waveform is an optical channel. See To Perform Dark-Level and User
Wavelength Gain Compensations on page 3--98.

H Optimize the vertical resolution before taking this measurement. See To Optimize the

Vertical Resolution on page B--69.