Incoming Inspection

1- 24

CSA8000B & TDS8000B User Manual

6.

Verify that the channel is operational: Confirm that the following statements
are true:

H The vertical scale readout for the channel under test shows a setting of

100 mV

, and a DC level is at about 2 divisions above center screen.

H The front-panel vertical POSITION knob (for the channel you are

testing) moves the DC level up and down the screen when rotated.
Return the DC level to 2 divisions above center screen before continuing.

H Turning the vertical SCALE knob to 50 mV changes the amplitude of

the DC level to about 4 divisions above center screen, and returning the
knob to 100 mV returns the amplitude to about 2 divisions above center
screen.

7.

Verify that the channel acquires in all acquisition modes: Push the
front-panel Acquisition MENU button to display the Acq setup dialog box.
Click each of the three acquisition modes, and confirm that the following
statements are true:

H Sample mode displays an actively acquiring waveform on-screen. (Note

that there is a small amount of noise present on the DC level).

H Average mode displays an actively acquiring waveform on-screen with

the noise reduced.

H Envelope mode displays an actively acquiring waveform on-screen with

the upper and lower extremes of the noise displayed.

8.

Close Acquisition setup dialog box: Push the Acquisition MENU button to
close the Acq setup dialog box.

9.

Verify the DC accuracy compensation: Do the following substeps:

a.

Select Measurement from the Setup menu. In the Meas Setup dialog box
that displays:

H Select as Source the channel under test. For example, select Main C3

for channel 3.

H Select Meas1.

H Set the Select Meas menu to Pulse > Amplitude > Mean.

b.

Push the Vertical MENU front-panel button to switch to the Vert Setup
dialog box.