Tektronix CSA8000B User Manual

Page 278

Advertising
background image

Appendix B: Automatic Measurements Reference

B- 16

CSA8000B & TDS8000B User Manual

Table B- 4: RZ Measurements - Amplitude (cont.)

Name

Definition

RZ Average Optical
Power (dBm)

The true average component of an optical signal, expressed in decibels. This measurement
results from the use of a hardware average power monitor circuit rather than from the
calculation of digitized waveform data.

Note: Average optical power measurements return valid results only on channels that contain
average power monitors. In general, all optical sampling module channels contain average
power monitors.

To determine RZ Average Optical Power (dBm), this measurement simply converts average
optical power (watts) to decibels using a log10 function referenced to 1mW. To determine
average optical power in watts, see the RZ Average Optical Power (Watts) measurement below.

For best average optical power measurement results:

H Use a factory-calibrated wavelength. If using the USER wavelength setting, ensure that it

is properly compensated by performing the User Wavelength Gain compensation found by
clicking the Optical button in the Vertical Setup dialog box.

H Compensate the optical channel, which corrects for minor DC variances in the average

power monitor as part of the compensation routine. To access, choose Compensation in
the Utilities menu of the application.

RZ Average Optical
Power (watts)

Where:

H DC Signal Current is the O/E-converter photo detector current in DC amps
H Conversion Gain is the O/E-converter photo detector gain in amps/watts

Note: Average optical power measurements return valid results only on channels that contain
average power monitors. In general, all optical sampling module channels contain average
power monitors.

To obtain accurate results, the O/E converter is calibrated at a fixed number of factory-calibrated
wavelengths to determine the conversion gain of the O/E converter at each wavelength.

For best average optical power measurement results:

H Use a factory-calibrated wavelength. If using the USER wavelength setting, ensure that it

is properly compensated by performing the User Wavelength Gain compensation found by
clicking the Optical button in the Vertical Setup dialog box.

H Compensate the optical channel, which corrects for minor DC variances in the average

power monitor as part of the compensation routine. To access, choose Compensation in
the Utilities menu of the application.

Average Optical Power (watts)

=

DC Signal Current (DC amps)

Conversion Gain (amps

ፒwatts)

Advertising
This manual is related to the following products: