Acquiring Waveforms

3- 24

CSA8000B & TDS8000B User Manual

Use the procedure that follows to set the data-acquisition mode and specify
acquisition start and stop methods. For more detailed information, display online
help when performing the procedure.

Overview

To set acquisitions modes

Control elements and resources

Prerequisites 1.

Instrument must be installed with sampling modules in
place before powering on the instrument. Instrument
must be powered up, with horizontal and vertical
controls setup. Triggering should also be set up.

See the sampling-module user manuals for sampling

module installation. See page 3--48 for trigger setup.

To select an

Acquisition mode

2.

Push the Acquisition MENU button to display the Acq
Setup dialog box.

Select the

Acquisition mode

Set a

sample count

3.

Click an option button to select the acquisition mode;
choose from the following modes:

H Sample
H Average
H Envelope

For Average mode only, enter the number of samples to
to average in the Average box.

To Set Acquisition Modes