Signal Connection and Scaling

This section presents an overview of the instrument features related to setting up
the input signal for digitizing and acquisition. It addresses the following topics:

H Where to find information for installing sampling modules and connecting

input signals

H How to turn on channels and adjust their vertical scale, position, and offset

H How to set the horizontal scale, position, and record length of the Main (time

base) View

NOTE. Terminology: This manual uses the terms vertical acquisition window and
horizontal acquisition window. These terms refer to the vertical and horizontal
range of the acquisition window, which defines the segment of the input signal
that the acquisition system acquires. The terms do not refer to any windows or
display windows on screen. See Conventions on page xii.

Vertical

position

Vertical

offset

Horizontal

scale

Horizontal

record

length

Acquisition

system

Sampling

module

Vertical

scale

Display

system

Horizontal

position

Figure 3- 1: Acquisition and display controls

Use signal conditioning and scaling controls to ensure the instrument acquires
the data that you want to display, measure, or otherwise process. To ensure the
best possible data for further processing, you do the following:

H Set vertical scale to adjust the waveform size on screen. You can set vertical

offset to shift the vertical acquisition window up or down on the signal to
capture the portion you want.

Why Use?