Installation

CSA8000B & TDS8000B User Manual

1- 13

Table 1- 1: Additional accessory connection information

Item

Description

Monitor

If you use a non-standard monitor, you may need to change the
the Windows display settings to achieve the proper resolution
for your monitor.

Printer

Connect the printer to the EPP (enhanced parallel port)
connector directly. If your printer has a DB-25 connector, use
the adapter cable that came with your printer to connect to the
EPP connector. For information on printer usage, see Printing
Waveforms on page 3-126.

Other

Refer to the Application release notes on your System Rebuild
CD for possible additional accessory installation information
not covered in this manual.

Power On the Instrument

Follow these steps to power on the instrument for the first time.

1.

Check that the line fuses are correct for your application. Both fuses must be
the same rating and type. Fuse types require a unique cap and fuseholder. See
Table 1--2 and Figure 1--4.

Table 1- 2: Line fuses

Fuse type

Rating

Fuse part number

Cap & fuseholder
part number

0.25 x 1.250 inch

8 A, fast blow, 250 V

159-0046-00

200-2264-00

5 x 20 mm

6.3 A, fast blow, 250 V

159-0381-00

200-2265-00

Fuses

AC power

Power switch

Figure 1- 4: Line fuse and power cord connector locations, rear panel