Tektronix CSA8000B User Manual

Page 253

Advertising
background image

Appendix A: Specifications

CSA8000B & TDS8000B User Manual

A- 3

Table A- 2:

System - Timebase

(cont.)

Description

Characteristics

n Time internal ac-
curacy, locked to in-
ternal 10 MHz refer-

2

Strobe placement accuracy for a given horizontal interval and position
on same strobe line per table below. Contribution from 80E04 sampling
module is included in specification.

0

ence mode

2

Range

Time Interval Accuracy

≤ 20 ps/div

1 ps + 1% of interval

≥ 21 ps/div

8 ps + 0.01% of interval

Horizontal deskew
range and resolution

--500 ps to +100 ns on any individual channel in 1 ps increments.

1

The total number of samples contained in a single acquired waveform record
(memory length in IEEE 1057, 2.2.1).

2

This is for

≤ 100 kHz trigger rate. The 80E04 sampling module is included in this

specification.

Table A- 3:

System - Trigger

Description

Characteristics

Trigger sources

External Direct Edge Trigger, External Prescaled Trigger, Internal Clock
Trigger, and Clock Recovery (with appropriately equipped optical
modules)

Auto/normal mode

Normal mode: wait for trigger

Auto mode: Trigger automatically generated after 100 ms time-out

Slope + or -- select

Edge + mode: Triggers on positive-slewing edge

Edge -- mode: Triggers on negative-slewing edge

High frequency on/off
select

High Frequency ON mode: Removes trigger hysteresis and improves
sensitivity. Should be used when trigger slew rate exceeds 1 V/ns.

High Frequency OFF mode: Retains trigger hysteresis and improves
noise rejection at low slew rates.

Metastability Reject
On/Off select

Metastability Reject On mode: Upon detection of trigger and holdoff
collision, time base will reject the sampled point.

Metastability Reject Off mode: Allows metastable points caused by
trigger/holdoff collisions to display.

Gated Trigger

±5 V maximum. See the Gated Trigger Input descriptions, on beginning
page A--8.

Variable trigger hold
off range and resolu-
tion

Adjustable 5 s to 50 ms in 0.5 ns increments. When External
Prescaled Trigger mode is used, holdoff period applies to the Prescaled
input divided by 8.