Tektronix CSA8000B User Manual

Page 50

Advertising
background image

Incoming Inspection

1- 32

CSA8000B & TDS8000B User Manual

8.

Verify that the Mag2 time base operates:

a.

Push the Mag1 button to remove the display of the Mag1 time base.

b.

Perform steps 6 and 7, but use the Mag2 button instead of the Mag1.

This test verifies that the Gated Trigger (GT Option) is functional. This test is
done using a front-panel signal and a rear-panel TTL connection.

Equipment
required

One 50

Ω BNC cable, such as Tektronix part number 174-1341-00

One SMA cable, such as Tektronix part number 174-1427-00

One 50

Ω terminator cap, such as Tektronix part number 011-0049-02.

One SMA 10X attenuator (20 dB attenuator), SMA connector, such as
Tektronix part number 015-1003-00

One electrical (80E00-series) sampling modules.

Prerequisites

This test applies only to instruments that include option GT.

1.

Initialize the instrument: Push the front-panel DEFAULT SETUP button,
and click Yes in the confirmation dialog box.

2.

Push the channel 3 button to select it. The button lights and the channel
display comes on. See Figure 1--15.

Channel

buttons

Figure 1- 15: Channel button location

Perform Gated Trigger

Test

Advertising
This manual is related to the following products: