3.1 Electrical connection

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

15

The following cables are available for these connections:

6.2143.200 Cable for one remote connection via

Remote line 1.

To connect a Liquid Handling Unit with a peristaltic pump, e.g.:

2.833.0010

IC Liquid Handling Pump Unit

2.833.0050

IC Liquid Handling Ultrafiltration Unit

6.2143.210 Cable for two remote connections via

Remote lines 1 and 2.

To connect a Liquid Handling Unit with a peristaltic pump and a

suppressor or sample preparation module, e.g.:

2.833.0020

IC Liquid Handling Suppressor Unit

2.833.0030

IC Liquid Handling Sample Preparation Unit

or

To connect two peristaltic pumps, e.g. for dialysis:

2.833.0010

IC Liquid Handling Pump Unit

2.833.0040

IC Liquid Handling Dialysis Unit

6.2143.220 Cable for three remote connections via

Remote lines 1, 2 and 4.

To connect up to three Liquid Handling Units, e.g. for dialysis with

suppression:

2x

2.833.0010

IC Liquid Handling Pump Unit

and 1x

2.833.0040

IC Liquid Handling Dialysis Unit

(three peristaltic pumps are switched on and off)

or

Ultrafiltration with suppression

1x

2.833.0020

IC Liquid Handling Suppressor Unit

and 1x

2.833.0050

IC Liquid Handling Ultrafiltration Unit

(two peristaltic pumps are switched on and off

and the sample preparation module is switched

on)

The remote line via which the control of the Liquid Handling Unit takes

place is determined by the wiring of the cable. The active remote line is

printed on the 833-end of the cable. This remote line must be set

accordingly in the control software «761 Compact IC» or «IC Net», see

Section 4.3.7 In the «761 Compact IC» Instructions for Use and Section

6.5.3 in the «IC Net» Instructions for Use.