9 Troubleshooting - maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

82

9.2.6

Suppressor module / Sample preparation module

The 1.753.0010 Suppressor Module and the 1.793.0010 sample

preparation module have the same technical construction so their

maintenance is described jointly. Wherever the different chemistry

requires a different treatment for the modules this is expressly

mentioned in the text.

9.2.6.1

Regenerating the actuator units

If the suppressor units of the suppressor module or the cation

exchanger units of the sample preparation module come into contact

with certain heavy metals (e.g. iron) or organic contaminants for a

longer period of time then these can no longer be completely removed

with the regeneration solutions normally used (50 mmol/L H

2

SO

4

or 50

mmol/L HClO

4

). This affects the capacity of the actuator units of the

module. If such capacity problems occur then the actuator units must

be treated as follows:

1 Disconnect suppressor module/sample preparation

module from IC system

Disconnect the module from column and detector.

2 Regenerate suppressor module

Rinse all 3 actuator units in sequence for approx. 10 min with

one of the following solutions:

Contamination with heavy metals

0.2 mol/L H

2

SO

4

+ 0.1 mol/L oxalic acid (for Fe

2+

/Fe

+3

)

Contamination with organic substances

0.2 mol/L H

2

SO

4

/ acetone

20%

Regenerate sample preparation module

Rinse all 3 actuator units in sequence for approx. 10 min with

one of the following solutions:

Contamination with heavy metals

0.2 mol/L HClO

4

+ 0.1 mol/L oxalic acid (for Fe

2+

/Fe

+3

)

Contamination with organic substances

0.2 mol/L HClO

4

/ acetone 20%

The 6.1826.050/6.1826.010 Pump tubing consists of PVC and must

therefore not be rinsed with solvents that contain acetone. In this case

you should either use different pump tubing or use a different pump

for rinsing.