7.5 Making the capillary connections

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

61

Immerse the other end of aspiration tubing 16 in the sample

container and fix it in position. Ultrapure water is normally

used as the acceptor solution; it must first be degassed (with

N

2

, He or vacuum).

3 Connect dialysis cell

Screw PTFE capillary 60) (6.1803.040; L =1 m) to PEEK

coupling 21 (6.2744.034) on the delivery side of the pump

tubing with the orange-yellow stoppers of channel 1 from

the 833 Dialysis Unit using a PEEK compression fitting 17

(6.2744.010) (see Fig. 7 and Fig. 21). Lead the other end of

the capillary through one of the feedthroughs 14 at the front

or the side of the 820 Separation Center in the column

compartment and screw it on to inlet opening 50 of upper

part 43 of the dialysis cell with a PVDF compression fitting 64

(6.2744.000) (see Fig. 15 and Fig. 21).

Screw PEEK capillary 61 (6.1831.050; L =40 cm)(see Fig. 21)

to outlet opening 49 of cell upper part 43 of the dialysis cell

using a PVDF compression fitting 64 (6.2744.000) (see Fig.

15).

Use only the above-mentioned

6.2744.000 PVDF compression

fittings for the dialysis cell connections. If 6.2744.010 PEEK

compression fittings are used then stress cracks could occur in the

dialysis cell !

4 Connect injection valve

Screw the PEEK capillary 61 (6.1831.050; L = 40 cm)

attached to outlet 49 to connection "1" of valve A 56 using a

6.2744.010 PEEK compression fitting.

Screw PEEK capillary 62 (6.1831.060; L =1 m) to connection

"2" of valve A 56 using a 6.2744.010 PEEK compression

fitting and lead it out of the column compartment through one

of the feedthroughs 14 on the front or side of the 820

Separation Center. Attach the other end of the PTFE capillary

62 to PEEK coupling 18 on the suction side of the orange-

yellow pump tubing of channel 2 from the 833 Dialsis Unit

using a PEEK compression fitting 17 (6.2744.010) (see Fig. 7

and Fig. 21).

5 Connect waste container

Screw PTFE capillary 63 (6.1803.040; L =1 m) (23 in Fig. 7)

to PEEK coupling 21 (6.2744.034) on the delivery side of the

pump tubing with the orange-yellow stoppers of channel 2

from the 833 Dialysis Unit using a PEEK compression fitting

17 (6.2744.010) (see Fig. 7 and Fig. 21). Lead the other end

of the capillary into a sufficiently large waste container and fix

it in position.