9.2 Care and maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

83

3 Connect suppressor module/sample preparation

module to IC system

Reconnect the module to the IC system. If there are still

capacity problems then the actuator rotor must be replaced

(see Section 9.2.6.3).

9.2.6.2

Cleaning the actuator module

Cleaning the suppressor module or sample preparation module may be

necessary in the following cases (see also Section 9.1):

Irreparable blockage of the suppressor

module/sample preparation module

(solutions can no longer be transported by the

suppressor)

Irreparable blockage of the actuator of the particular

module

(actuator can no longer be switched to next position)

Clean the connection piece and actuator rotor as follows (see Fig. 26):

1 Disconnect suppressor module/sample preparation

module from IC system

Disconnect the inlet of the suppressor module/sample

preparation module from the column and detector outlet.

2 Dismantle actuator module

Unscrew sleeve nut 85 from actuator holder 88.

Pull connection piece 86 and actuator rotor 87 out of actuator

holder 88.

Loosen connection piece 86 from actuator rotor 87.

3 Clean inlets and outlets

Connect each of the 6 pieces of capillary tubing attached to

connection piece 86 to the pump in sequence and pump

ultrapure water through them.

Check whether any liquid is escaping from connection piece

86. If one of the inlets or outlets remains blocked then the

connection piece must be replaced.

Order number 6.2832.010 for suppressor module

6.2835.010 for sample preparation module

4 Clean actuator rotor

Clean the sealing surface of actuator rotor 87 using a lint-free

cloth and ethanol.