4 assembling the dialysis cell, Assembling the dialysis cell – Metrohm 833 IC Liquid Handling Unit User Manual

Page 64

Advertising
background image

7 833 IC Liquid Handling Dialysis Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

56

7.4

Assembling the dialysis cell

This section describes the initial assembly of the dialysis cell. Before it

is used a dialysis membrane must be inserted. This is done as follows

(see Fig. 15):

1 Prepare dialysis cell

Remove dialysis cell 42 (6.2729.100) from the packaging and

screw out the four 6.2744.060 Blank stoppers .

Use the 6.2621.070 Allen key to loosen the 5 screws 47

completely, remove lower part 46 from upper part 43 and

remove sealing ring 44.

2 Clean dialysis cell

Thoroughly rinse sealing ring 44, upper part 43 and lower

part 46 of the dialysis cell with ultrapure water and dry with N

2

or a lint-free tissue.

Only ultrapure water or ethanol should be used for cleaning the

dialysis cell, other organic solvents (e.g. acetone) will damage the

Plexiglas cell!

3 Prepare dialysis membrane

Use the 6.2831.010 Tweezers to remove a new dialysis

membrane 45 (e.g. 6.2714.010) from the packaging and

immerse in a Petri dish filled with ultrapure water for approx. 2

min until the membrane is thoroughly soaked with water.

4 Insert dialysis membrane

Place upper part 43 on a paper tissue with its inner side

uppermost.

Place sealing ring 44 in the recess provided for it in upper

part 43.

Use the 6.2831.010 Tweezers to place the wet dialysis

membrane 45 (e.g. 6.2714.010) inside sealing ring 44 on

upper part 43.

Make sure that the dialysis membrane soaked with water does not dry

out before insertion as otherwise it cannot be used !

5 Close dialysis cell

Place lower part 46 on upper part 43 so that the two parts are

properly aligned with each other.

Use the 6.2621.070 Allen key to screw in the 5 screws 47

completely and tighten them up well.