2 settings in ic net, Settings in ic net – Metrohm 833 IC Liquid Handling Unit User Manual

Page 55

Advertising
background image

6.2 Connection to modular IC system for cation separation

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

47

Immerse the other end of the aspiration tubing in a container

with rinsing solution (normally deionized water) and fix it

position.

Pull outlet capillary 39 (marked with “Waste”) at sample

preparation module connection 34 (see Fig. 13) through one

of the openings 18 in the rear panel of the 820 IC Separation

Center.

Lead outlet capillary 39 into a sufficiently large waste bottle

and fix it in position.

8 Start-up 833 IC Liquid Handling Sample Preparation

Unit

Switch on 833 LH Pump Unit with mains switch 8.

Set the contact pressure for both cassettes: press contact

lever 5 until the solutions are just aspirated. Then press the

contact lever up by 1 further click to achieve the optimum

contact pressure.

Check all the tubing from the storage containers to the 833 IC

Liquid Handling Unit and the sample preparation module

through to the waste bottles for leaks. If any liquid should be

leaking then the corresponding connection must be tightened

up or replaced.

Before the sample preparation module is used for the first

time the cation exchanger material must be conditioned. This

is done by injecting e.g. 30% NaOH 15 times (5 injections per

unit). You can use the program described in the following

section for the system “Neutralisation.smt” with a running

time of 20 min; the column does not need to be mounted.


6.2.2

Settings in IC Net

When setting up the 833 IC Liquid Handling Sample Preparation Unit

and the 833 IC Liquid Handling Unit the “New system wizard” of the IC

Net program demands the assignment of an “Event line” for the

Actuator, i.e. the sample preparation module. Although this does not

exist you should still select any unused Event line so that the “New

system wizard” can continue. In fact the sample preparation module is

controlled from the 766 IC Sample Processor (see Fig. 14).

The signal for switching the internal suppressor module of the 820 to

the next position is not controlled by the program of the 766 IC Sample

Changer. Instead the parameter “Auto Step” is set to “fill” in the

parameter window of the 820 IC Separation Center in IC Net (double-

click on 820 symbol).

Advertising