7 dialysis cell, Dialysis cell, Fig. 27 – Metrohm 833 IC Liquid Handling Unit User Manual

Page 96: Adjusting the actuator step, Fig. 27 : in one of

Advertising
background image

9 Troubleshooting - maintenance

830 IC Interface / Gebrauchsanweisung 8.830.1001

88

3 Adjust actuator step

Turn setting screw 12 as shown in Fig. 27 a little in the

required direction so that the rotor comes to rest in a valid

position.

Activate <STEP> in the software.

Check the position of the rotor in all 3 possible settings as

shown in Fig. 27.

Repeat the setting procedure until the rotor is in a valid

position in all of the 3 possible settings. Always wait for at

least 10 s before triggering the next STEP command.


optimal

zulässig

zulässig

unzulässig

unzulässig

Fig. 27:

Adjusting the actuator step


9.2.7 Dialysis

cell

The dialysis membrane is very sensitive to contact and drying out, so

avoid touching it with your fingers and ensure that the membrane never

dries out when inserted in the cell.
The working life of a dialysis membrane is approx. 1 week if used

frequently. It should always be exchanged if there are any indications

that the yield has become reduced or that it is being blocked.
Follow the instructions for the installation of the dialysis cell (Section 7.4

and 7.5), its shut down (Section 9.2.3) and for replacement of the

dialysis membrane (Section 9.2.7.1).

Advertising