background image

4.1 Connection of the suppressor module

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

27

4 833 IC Liquid

Handling Pump Unit

This version is mainly used for supplying a Metrohm Suppressor

Module MSM built into a 2.820.0X30 IC Separation Center with

regeneration and rinsing solutions. The Metrohm Suppressor Module

MSM is described in detail in Section 2.9.7 of the 819/820 Instructions

for Use.

4.1

Connection of the suppressor module

In order to protect the suppressor module against foreign particles an

inline filter is mounted between the 833 IC Liquid Handling Unit and

the inlet capillary of the suppressor module, see Fig. 7. The most

suitable unit is the 6.2821.120 Filter unit PEEK included in the

standard equipment.
The suppressor module is connected to the 833 IC Liquid Handling

Pump Unit in accordance with Section 2.9.7 of the 819/820 Instructions

for Use. The most important points are described again below, with

underlined numbers ## referring to the parts and controls in the

819/820 Instructions for Use:


Fig. 8:

Connections to suppressor module of 820 Separation Center

64 Suppressor inlet

capillary for eluent

65 Suppressor inlet

capillary for H

2

SO

4

66 Suppressor outlet

capillary for H

2

SO

4

67 Suppressor outlet

capillary for H

2

O

68 Suppressor inlet

capillary for H

2

O

69 Suppressor outlet

capillary for eluent

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4

1

3

2

67

64

65

66

68

69