9.2 Care and maintenance

830 IC Interface / Gebrauchsanweisung 8.830.1001

85

8 Connect and condition suppressor module

Reconnect suppressor module to the IC system (see Section

5.1).

Before moving the suppressor to the next position for the first

time rinse the 3 actuator units with solution for 5 min.

Connect and condition sample preparation module

Reconnect sample preparation module to the IC system (see

Section 6.1 and Section 6.2).

Before moving the suppressor to the next position for the first

time rinse the 3 actuator units with solution for 5 min.

9.2.6.3 Replacing

the actuator rotor

It may be necessary to replace actuator rotor 87 contained in the cation

exchanger of the suppressor module or sample preparation module in

the following cases (see also Section 9.1):

Irreparable loss of exchanger capacity

(recognizable at suppressor by reduced phosphate

sensitivity and/or large increase in baseline)

Irreparable blockage of the suppressor

(solutions can no longer be transported by the

suppressor)

As well as replacing actuator rotor 87 (6.2832.000 in suppressor

module, 6.2835.000 in sample preparation module), it may also be

necessary to replace connection piece 86 (6.2832.010 in suppressor

module, 6.2835.010 in sample preparation module). Exchange these

components in the following way (see Fig. 26):

1 Disconnect suppressor module/sample preparation

module from IC system

Disconnect the inlet of the suppressor module/sample

preparation module from the column and detector outlet.

2 Dismantle actuator module

Unscrew sleeve nut 85 from actuator holder 88.

Pull connection piece 86 and actuator rotor 87 out of actuator

holder 88.

Loosen connection piece 86 from actuator rotor 87.

3 Clean actuator rotor

Clean the sealing surface of the new actuator rotor 87 using

a lint-free cloth and ethanol.