7 833 IC Liquid Handling Dialysis Unit
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
54
7.3
Settings in IC Net
7.3.1 Dialysis
without suppression
The program steps required for dialysis are all controlled by the time
program of the 819 IC Detector.
Fig. 18:
Time program for dialysis without suppression
Make sure that the remote control of the peristaltic pumps
833 Pump
and
833 Dialysis (see Section 7.2.1) via the 6.2143.210 cable (see
Section 3.1.2) corresponds to the remote configuration of the detector
program.
Rinse time
Dialysis time
Transfer time
Start of 833 Pump (Sample)
Start of 833 Pump (Acceptor)
Stop of acceptor solution
Trigger auto-zero at 819 Detector
Fill sample loop
Transfer to sample loop
Injection (Start evaluation)
Stop of 833 Pump (Acceptor)
Stop of 833 Pump (Sample)