Fig. 24, Tubing connection system, Also fig. 24 ) – Metrohm 833 IC Liquid Handling Unit User Manual

Page 79

Advertising
background image

8.3 Connect filtration cell

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

71

3 Sample feed to filtration cell

Lead the other end of PTFE capillary 77 through one of the

openings at the front of the 820 IC Separation Center into the

column compartment.

Attach this end of the PTFE capillary to inlet opening 73 on

the base of the ultrafiltration cell using a PVDF compression

fitting 64 (see Fig. 22).

4 Tubing connection filtration cell – waste

Attach one end of PTFE capillary 78 to outlet opening 74 on

the base of the ultrafiltration cell with a PVDF compression

fitting 64 (see Fig. 22).

Lead the other end of PTFE capillary 77 out through opening

18 on the rear panel of the 820 IC Separation Center and into

a waste bottle

(e.g. optional: 6.1608.070 Bottle with 6.1602.150 GL45 bottle

attachment).

Fig. 24:

Tubing connection system

16 6.1803.080 PTFE capillary

for sample solution feed

17 6.2744.010 PEEK compression

fitting

18 6.2744.034 Coupling

21 6.2744.160 PEEK coupling with

tubing security device

64 6.2744.000 PVDF compression

fitting

66 6.2729.110 Filtration cell

Filtrat

833

820

A

79

Column

Sample

Waste

80

16

78

64

66

77

64

18

17

75

81

Eluent

21

Waste

76