2 connecting to 819 ic detector, Connecting to 819 ic detector, Fig. 5 – Metrohm 833 IC Liquid Handling Unit User Manual

Page 24

Advertising
background image

3 Basic instrument

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

16

3.1.2

Connecting to 819 IC Detector


Connect the remote connection of the 819 IC Detector to the necessary

remote connections 14 and 15 on the 833 IC Liquid Handling Unit. The

peristaltic pump of the Liquid Handling Unit is controlled via remote

connection Remote Pump 15, this must always be connected. The

connection Remote Actuator 14 is only required for the 833 IC Liquid

Handling Suppressor Unit and the 833 IC Liquid Handling Sample

Preparation Unit; it is used to switch the suppressor or sample

preparation module. The particular modules are connected to the

socket Module 13.

Fig. 5:

Connecting the 833 IC Liquid Handling Unit to 819 IC Detector

The following cables are available for these connections:

6.2143.200 Cable for one remote connection via

Remote line 1.

To connect a Liquid Handling Unit with a peristaltic pump, e.g.:

2.833.0010

IC Liquid Handling Pump Unit

2.833.0050

IC Liquid Handling Ultrafiltration Unit

6.2143.210 Cable for two remote connections via

Remote lines 1 and 2.

To connect a Liquid Handling Unit with a peristaltic pump and a

suppressor or sample preparation module, e.g.:

2.833.0020

IC Liquid Handling Suppressor Unit

2.833.0030

IC Liquid Handling Sample Preparation Unit

or

To connect two peristaltic pumps, e.g. for dialysis:

2.833.0010

IC Liquid Handling Pump Unit

2.833.0040

IC Liquid Handling Dialysis Unit

6.2143.210

816

833

6.2143.200

816

833

Cable
to
Module

819

819

Advertising