6.1 Connection to modular IC system for neutralization

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

41

8 Start-up 833 IC Liquid Handling Sample Preparation

Unit

Switch on 833 LH Pump Unit with mains switch 8.

Set the contact pressure for both cassettes: press contact

lever 5 until the solutions are just aspirated. Then press the

contact lever up by 1 further click to achieve the optimum

contact pressure.

Check all the tubing from the storage containers to the 833 IC

Liquid Handling Unit and the sample preparation module

through to the waste bottles for leaks. If any liquid should be

leaking then the corresponding connection must be tightened

up or replaced.

Before the sample preparation module is used for the first

time the cation exchanger material must be conditioned. This

is done by injecting e.g. 30% NaOH 15 times (5 injections per

unit). You can use the program described in the following

section for the system “Neutralisation.smt” with a running

time of 20 min; the column does not need to be mounted.

6.1.2.2

Settings in IC Net

When setting up the 833 IC Liquid Handling Sample Preparation Unit

and the 833 IC Liquid Handling Unit the “New system wizard” of the IC

Net program demands the assignment of an “Event line” for the sample

preparation module and the suppressor module. Although this does not

exist you should still select any unused Event line so that the “New

system wizard” can continue. In fact the sample preparation module is

controlled from the 766 IC Sample Processor and the suppressor

module from the Event Line of the 820 valve B (see Fig. 12).
Apart from the main program “Neutralisation.smt” two further programs

should be created for the 766 IC Sample Processor in separate system

files. The 766 program in the system file “Start-Neutralisation.smt” only

starts the peristaltic pump (833 Pump) and rinses the sample

preparation module in all three positions. “End-Neutralisation.smt”

switches the pump off (833 Pump).
The peristaltic pumps of the 833 IC Liquid Handling Sample Preparation

Unit (833 Sample Prep) and the 833 IC Liquid Handling Suppressor

Unit (833 Suppressor, see Fig. 12) start immediately when the system

is activated (recording the baseline). This is done by activating the

option ’Start pump with startup hardware’ in IC Net for each

instrument.