5.2 Handling the suppressor module

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

35

5.2

Handling the suppressor module

General

The Metrohm Suppressor Module MSM of the 833 IC Liquid

Handling Suppressor Unit consists of a total of three suppressor

units, which are used in sequence for suppression, regenerated with

sulfuric acid and rinsed with water. In order to be able to record each

chromatogram under comparable conditions, work is normally carried

out with a freshly regenerated suppressor. Switching takes place either

automatically together with the valve switching or manually.

Correct connection

The three suppressor inlets and outlets numbered 1

3 each have two

permanently attached PTFE capillaries that must be connected as

described in Section 5.1 (see Fig. 10).

Flow direction

The suppressor units must never be regenerated with H

2

SO

4

in the

same direction as which the eluent is transported. This is why the inlet

and outlet capillaries must always be mounted as described in

Section 5.1 and shown in Fig. 10.

Never switch to the next position when dry

The suppressor module must never be switched to the next position

when it is dry, as there is a risk that it could be blocked.

No recycling

The recycling method (return of the eluent to the storage container)

must not be used with the suppressor.