9 Troubleshooting - maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

76

9 Troubleshooting -

maintenance

9.1

Faults and their remedies

If difficulties occur when carrying out analyses with the IC system, then

it is advisable to search for their causes in the following sequence

column

pump

eluent

IC system. An overview of faults, their

possible causes and remedies can be found in the Instruction for Use

of both the 761 Compact IC (Section 5.3) and the 819/820 Modular IC

System (Section 4.3.2).

In addition to these general faults, the following table lists those faults

that could occur during the operation of an IC Liquid Handling Unit.

Fault

Cause

Remedy

Inadequate or no supply
at all from pump

Contact pressure too low




Pump tubing faulty

6.2821.120 PEEK filter unit
blocked

Connection blocked





Suppressor module /
Sample preparation module

Actuator rotor blocked

Set contact pressure correctly: first press
contact lever 5 on the tubing cassette
down as far as it will go, then up again
until the solution is just aspirated. Then
press the contact lever up by 1 click to
achieve the optimal contact pressure.

Replace pump tubing
(see Section 9.2.4)

Replace 6.2821.130 filter (see
connection description, Section 9.2.5)

Check tubing and connections step by
step and clean or replace as necessary. If
the suppressor inlets or outlets are
blocked then dismantle the suppressor
and clean it (see Section 9.2.6.2).


Replace actuator rotor
(see Section 9.2.6.3).