Metrohm 833 IC Liquid Handling Unit User Manual

Page 50

Advertising
background image

6 833 IC Liquid Handling Sample Preparation Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

42

Neutralisation.smt
1. Program input for the 766 IC Sample Processor:

001 Ctrl INIT

819

002 Move

sample

003 Lift work

004

Ctrl

FILL B / STEP 1

005 Pump

120

s

006

Ctrl

FILL A 1

007 Ctrl INJECT

B1

008 Wait 60

s


009 Ctrl ZERO

1

010 Ctrl INJECT

A1

011

Ctrl

STEP MSM 833

Initializes remote lines at 819

Moves needle to sample position

Moves lift with needle to working position

Switches injection valve B at 820 to "Fill"

Transfers sample to first sample loop

Switches injection valve A at 820 to "Fill"

Switches injection valve B at 820 to "Inject"

Transfers sample via sample preparation module to
enrichment column (*)

Triggers autozero at 819 IC Detector

Switches injection valve A at 820 to "Inject"

Switches sample preparation module to next position

(*) The time for transferring the sample via the sample preparation module must

be adequate, but should be kept as short as possible in order to minimize
contamination of the downstream enrichment column with unwanted anions
from the water used. This is also the reason why the blank value of this water
should be measured regularly for calibration purposes.

Start-Neutralisation.smt
2. Save the system “Neutralisation.smt” under a new name “Start-

Neutralisation.smt”.

3. Remove the data recorder by clicking on the recorder icon with the

right-hand mouse key and then selecting “Unlink”.

4. Change the program for 'Start-Neutralisation.smt' as follows and

save it again:

001 Ctrl INIT

819

002

Ctrl

PUMP 833 on

003 Wait 300

s

004

Ctrl

STEP MSM 833

005 Wait 300

s

006

Ctrl

STEP MSM 833

007 Wait 300

s

Initializes remote lines at 819

Switches on 833 Pump Unit 833

Rinsing time

Switches sample preparation module to next position

Rinsing time

Switches sample preparation module to next position

Rinsing time

End-Neutralisation.smt
5. Save the system “Start-Neutralisation.smt” under a new name “End-

Neutralisation.smt”.

6. Change the program for “End-Neutralisation.smt” as follows and

save it again:

001 Ctrl INIT

819

002

Ctrl

PUMP 833 off

Initializes remote lines at 819

Switches off 833 Pump Unit 833

Advertising