Ction 9.2.7.1 and, Section 9.2.7.1 ), Ion 9.2.7.1) – Metrohm 833 IC Liquid Handling Unit User Manual

Page 97

Advertising
background image

9.2 Care and maintenance

830 IC Interface / Gebrauchsanweisung 8.830.1001

89

9.2.7.1

Replacing the dialysis membrane

It may be necessary to replace the dialysis membrane in the following

circumstances (see also Section 9.1):

Reduced yield after dialysis

Membrane damaged by drying out or deposits or

bacterial growth

Irreparable blockage of sample channel (sample can

no longer be transported through dialysis cell)

Proceed as follows when replacing the membrane (see Fig. 15):

1 Dismantle dialysis cell

Screw off the four inlets and outlets of dialysis cell 42 and

take the cell out of the cell holder.

Use the 6.2621.070 Allen key to loosen the 5 screws 47 and

separate lower part 46 from upper part 43

Remove old dialysis membrane 45.

2 Clean dialysis cell, insert dialysis membrane and

reassemble cell

See Steps 2-6 Section 7.4 Assembling the dialysis cell.

3 Connect dialysis cell

See dialysis cell Section 7.5.2 Connection of the dialysis

cell.

4 Condition dialysis membrane

See also Section 7.5.2.

Immerse the 2 pieces of aspiration tubing 16 (see Fig. 21) in

the acceptor solution (degassed ultrapure water).

Switch on the peristaltic pump in «IC Net».

Rinse the dialysis system for approx. 20 min with ultrapure

water. Check that the same amount of solution emerges from

both the waste outlets in the waste bottle.

Check all tubing from the storage containers to the tubing

cassettes and dialysis cell and up to the waste bottle for

leaks. If any leaks are found then the corresponding

connection must either be tightened up or replaced.

If any air bubbles remain in the dialysis cell then screw off

PEEK tubing 61 (acceptor solution) and PTFE tubing 58

(sample) from outlets 49 and 52 of the dialysis cell and wait

until the air bubbles have vanished. Then screw the tubing

back on again.

Switch off peristaltic pump.

Advertising