7 dialysis procedure, Dialysis procedure – Metrohm 833 IC Liquid Handling Unit User Manual

Page 74

Advertising
background image

7 833 IC Liquid Handling Dialysis Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

66

7.7 Dialysis

procedure

For determinations of samples by using dialysis it is advisable to

proceed in the following sequence:

1 Prepare acceptor solution

To avoid interference from air bubbles in the acceptor

channel the ultrapure water used for the acceptor solution

should always be degassed for at least 10 min by vacuum,

N

2

or He.

2 Prepare sample

In order to avoid blocking the sample channel, samples

containing a large amount of suspended or solid particles

should always be centrifuged using a benchtop centrifuge

at 10'000 min

-1

for 5 min.

3 Start up IC system

Switch on all instruments.

Start the IC Net software and load/create the required system

(see «IC Net» Instructions for Use).

Condition the IC column (see 819/820 Instructions for Use).

4 Start up dialysis system

Immerse both pieces of aspiration tubing 16 in the acceptor

solution.

Start the peristaltic pumps in «IC Net» by setting the remote

line 1 (833 Pump Unit) and remote line 2 (833 Dialysis Unit) in

the window of the 819 Detector.

Rinse the dialysis system with acceptor solution for approx.

10 min and then switch off the pump again.

5 Calibration

Immerse sample aspiration tubing 16 in the standard

solution.

Start the determination in IC Net and wait until the standard

chromatogram has been evaluated and the next standard

can be measured.

6 Sample determination

Immerse sample aspiration tubing 16 in the sample solution.

Start the determination in IC Net and wait until the

chromatogram has been evaluated.

Advertising