9 Troubleshooting - maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

80

Dialysis cell

To shut down the dialysis cell the acceptor and sample channels must

be rinsed with ultrapure water for approx. 10 min. The inlet and outlet

tubing is then screwed off from the dialysis cell and openings 49, 50,

51 and 52 (see Fig. 15) are each closed with a 6.2744.060 Blank

stopper.
If the dialysis system is not to be used for a long time then the dialysis

membrane must be removed and the cell has to be cleaned, see

Section 9.2.7.1 and Fig. 15.

Filtration cell

If the filtration cell is not to be used for a long time then must be rinsed

with ultrapure water for approx. 10 min. The inlet and outlet tubing is

then screwed off from the filtration cell and openings 71, 73 and 74

(see Fig. 22) are each closed with a 6.2744.060 Blank stopper.

9.2.4

Replacing the pump tubing

The pump tubing is a consumable with a limited working life, which is

why it should be replaced at regular intervals (approx. every 2 weeks for

continuous use).
The working life of the pump tubing depends primarily on the contact

pressure. This is why you should set the contact pressure correctly as

described in Section 3.2 and lift up the tubing cassettes completely

when the pump is switched off for a long time releasing spring lever 7

on the right-hand side (this means that the optimally set contact

pressure is retained).
As the pump always operates with the same direction of rotation this

means that the supplied 6.1826.0X0 Pump tubing can be used on both

sides. The pump tubing is replaced as follows:

1 Remove old pump tubing

Press contact lever 5 on the tubing cassette down as far as it

will go.

Loosen tubing cassette 4 from holding clip 6 by pressing

down spring lever 7 and unhinge it from the holding pin 2 on

the 833 IC Liquid Handling Unit (see Fig. 1).

Remove the old pump tubing.

2 Insert new pump tubing

Insert new pump tubing 19 (6.1826.0X0) in the tubing

cassette as shown in Fig. 7. Stopper 20 must click into

position in the corresponding holder on the left-hand side of

the tubing cassette.

Hinge the tubing cassettes on holding pin and press down

on the right-hand side until spring lever clicks into position in

holding clip 6. Take care that the pump tubing does not kink.