3 filter membrane selection, Filter membrane selection – Metrohm 833 IC Liquid Handling Unit User Manual

Page 83

Advertising
background image

8.4 Filtration

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

75

stable rate for many samples and then drop sharply, with a different

sample matrix it may diminish slowly and continuously.
The time when the filter membrane should be replaced ultimately

depends on the sample matrix and the specifications of the analytical

method used. In our experience very fine and suspended particles in

the sample matrix will block the filter membrane more rapidly than

coarse particles, which tend to be carried past the membrane in the

sample flow.
Instructions for replacing the filter membrane are given in

Section 9.2.8.1.

8.4.3

Filter membrane selection

You can apply existing sample preparation rules to filtration with the 833

IC Liquid Handling Ultrafiltration Unit. If you wish to use a different filter

membrane from that supplied as standard then please note that, even if

the particle size is known, the selection of a membrane with an

appropriate pore size does not automatically produce the required

results.
Some investigations have shown that the retention capability of normal

filter membranes does not always correspond to their specified pore

size. The following table shows the qualitative filtering effect of filter

membranes with different nominal pore sizes. The tests were carried

out on aqueous solutions containing silica particles with particles sizes

1.5 µm and 5 µm.

Test solutions:

silica particles in water

Pore size

of filter membrane

1

Effect

0.5%, 5 µm

0.15 µm

no breakthrough

0.5%, 5 µm

3 µm

no breakthrough

0.5%, 5 µm

8 µm

no breakthrough

0.5%, 5 µm

10 µm

breakthrough

2

0.5%, 5 µm

12 µm

no breakthrough

0.5%,

1.5

µm

0.15 µm

no breakthrough

0.5%,

1.5

µm

3 µm

breakthrough

(1)

Nominal pore size according to information supplied by manufacturer.

(2)

All the membranes came from a single manufacturer except this one.

Please also note that, because of their lack of thickness, the retention

capability of filter membranes may be lower than that of thicker filters

with the same nominal pore size. This should be taken into account

when choosing a suitable filter membrane.

Advertising