3.2 Assembly the pump tubing

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

19

3.2

Assembly the pump tubing

Pump tubing is a consumable whose working life depends on the

contact pressure. This is why when the pump is switched off for a long

time, the tubing cassette should be lifted by fully loosening spring

lever

7

on the right-hand side (in this way the correctly set optimal

contact pressure is retained).

The 6.1826.0X0 Pump tubing consists of PVC or PP and must

therefore not be rinsed with solvents that contain acetone. In this case

you should either use different pump tubing or use a different pump

for rinsing.

Fig. 7:

Assembly the pump tubing

4 Tubing cassette

5 Contact lever

7 Spring lever

16 6.1803.0X0 Aspiration

tubeing/capillary

PTFE tubing/capillary, depends on

version, see Section 10.2 Standard

equipment

17 6.2744.010 PEEK compression

fitting

18 6.2744.034 PEEK coupling

19 6.1826.0X0 Pump tubing

various tubing is available, depending

on the application, see Tab. 1.

20 Stopper

The color of the stopper indicates the

material and dimensions of the pump

tubing, see Tab. 1.

21 6.2744.160 PEEK Coupling

with tubing security device

22 6.2821.120 PEEK Filter unit

not contained in all versions, see

Section 10.2 Standard equipment

23 Discharge capillary

delivery side for connection to:

suppressor module

sample preparation module

dialysis cell

ultrafiltration cell

18

17

16 4

5

19

7

20 22

21

20 17

23

17

17

16

Advertising