6.3 Handling the sample preparation module

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

49

6.3 Handling

the

sample preparation module

General

The sample preparation module of the 833 IC Liquid Handling

Sample Preparation Unit contains a total of three cartridges, which

are used in sequence for exchanging cations, for regeneration, and for

rinsing with water. In order to be able to record each chromatogram

under comparable conditions work is normally carried out with a freshly

regenerated cartridge. Switching takes place either automatically

together with the valve switching or manually.

Correct connection

The three actuator unit inlets and outlets numbered 1, 2, 3 each have

two permanently attached PTFE capillaries that must be connected as

described in Section 6.1 and 6.2 (see Fig. 13).

Flow direction

The cation exchanger units of the sample preparation module must

never be regenerated in the same flow direction as that in which the

sample is transported. This is why the inlet and outlet capillaries must

always be mounted as described in Section 6.1 and 6.2 and shown in

Fig. 13.

Never switch to the next position when dry

The sample preparation module must never be switched to the next

position when it is dry, as there is a risk that it could be blocked.

No recycling

The recycling method (return of the eluent to the storage container)

must not be used with the sample preparation module.