Connect pump tubing for sample, 75 w, 75 using a – Metrohm 833 IC Liquid Handling Unit User Manual

Page 80: 78 to outlet opening, Lead the other end of ptfe capillary 77 out

Advertising
background image

8 833 IC Liquid Handling Ultrafiltration Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

72

75 6.1826.070 Pump tubing

(19 in Fig. 7) with yellow-yellow

stoppers for transferring the sample

76 6.1826.030 Pump tubing

(19 in Fig. 7)with orange-yellow

stoppers for transferring the filtrate

77 6.1803.070 PTFE capillary

(23 in Fig. 7)

connection pump tubing – filtration cell

(inlet)

for transferring the sample

78 6.1803.080 PTFE capillary

connection filtration cell – waste

79 6.1803.050 PTFE capillary

(23 in Fig. 7)

connection filtration cell (outlet) – pump

tubing

for transferring the filtrate

80 6.1831.100 PTFE capillary

connection pump tubing – valve

for transferring the filtrate

81 6.1831.040 PTFE capillary

connection pump tubing – valve

for transferring the filtrate

Use only the above-mentioned

6.2744.000 PVDF compression

fittings for the ultrafiltration cell connections. If 6.2744.010 PEEK

compression fittings are used then stress cracks could occur in the

ultrafiltration cell !

5 Transferring the filtrate

Connection filtration cell – pump tubing

Attach one end of PTFE capillary 79 to the outlet opening for

the filtrate 71 on the top of the ultrafiltration cell using a PVDF

compression fitting 64 (see Fig. 22) and lead out the other

end through one of the openings on the front of the 820 IC

Separation Center.

Screw this end onto PEEK coupling 18 on the suction side of

pump tubing 76 with the orange-yellow stoppers

(6.1826.030) using a PEEK compression fitting 17

(6.2744.010) (see Fig. 7, Fig. 24).

6 Transferring the filtrate

Connection pump tubing – injection valve

Screw one end of PEEK capillary 80 onto PEEK coupling 21

(6.2744.034) at the delivery side of pump tubing 76 with the

orange-yellow stoppers (6.1826.030) using a PEEK

compression fitting 17 (6.2744.010) (see Fig. 7, Fig. 24).

Lead PEEK capillary 80 through one of the openings 14 on

the front or side of the 820 IC Separation Center into the

column compartment and screw it onto connection “1” of the

injection valve instead of PTFE-aspiration tubing 63 (see

Fig. 21 and Fig. 23 of the 819/820 Instructions for Use) with a

PEEK compression fitting 17 (6.2744.010).

Tighten up the rotary nipple on the inside of connections 22

and 28 to fix PEEK capillary 80 in position.