Editing the Ultrasonic Device

Edit the ultrasonic device as follows, and select OK when complete:

Edit Ultrasonic Device Properties dialog

Use Defaults Select this checkbox to apply the default
parameters to the ultrasonic device. Deselecting this option
enables the parameter fields for editing.

Physical Offset Enter the physical distance between the face
of the ultrasonic depth sensor and the top (crown) of the pipe.

Scrubbing Window Enter the allowable difference between
individual ultrasonic pair readings and the average reading of
all the pairs. If any single pair exceeds the difference, the pair
will be flagged and removed from the calculation. The
remaining pairs will be averaged to calculate the depth
readings.

Temperature Offset This field should not require
modification for the FlowShark monitor. No additional
temperature compensation is necessary.